Step 3-6: Verify calibration points
[Purpose]
Verify that the grid pattern was detected correctly.
If there are any wrongly detected circles (calibration points), delete them.

[Procedure]
PC
Press '5 ®Õ¥¿ÂI' in the navigation area.
In the image in the left side of the screen, the positions with the green '+' marks indicate the detected circle.
PC
If '+' marks appear in positions other than the circles, use the procedure below to delete those detection results.
  1. Confirm the '+' mark number that appears on a position other than circles.
  2. In the table in the setup items area, click the lines corresponding to the number.
  3. Press the '§R°£' button below the table.
Additional
By clicking the "-°¾®t-" column header in the table on the setup items screen twice, calibration points can be sorted in descending order of "-°¾®t-" value. It makes easy to find wrongly detected calibration points.